Specification:
Measurement parameter Voltage mV | ||
Measuring range | Resolution | Accuracy |
0 ... 250 mV | 0.01 mV | ± 0.02% of reading + 2 Dgt |
Voltage DC V | ||
0 ... 30V | 0.001V | ± 0.02% of reading + 2 Dgt |
Current DC mA | ||
0 ... 24-mA | 0.001-mA | ± 0.02% of reading + 2 Dgt |
Simulation parameters | ||
Voltage mV | ||
0 ... 250 mV | 0.01 mV | ± 0.02% of reading + 2 Dgt |
Voltage DC V | ||
0 ... 12V | 0.001V | ± 0.02% of reading + 2 Dgt |
Current DC mA | ||
0 ... 24-mA | 0.001-mA | ± 0.02% of reading + 2 Dgt |
General Specifications PCE-LOC 20 | ||
Display modes | Measurement: mA / V / / mV Simulation: mA / V / / mV | |
Maximum input voltage | 30V DC | |
Input impedance measurement | mV / V:> 1 MΩ Current measurement: 10 Ω | |
Response time | <100 ms | |
Load impedance | > 10 kΩ at mV / V <750Ω at mA | |
Refresh rate display | 10 Hz | |
Isolation | 500V DC | |
Data storage | Internal memory 150000 readings | |
Interface | USB 2.0 | |
Display | 2.4" TFT LCD 240 x 320 pixels LED illuminated | |
Output voltage current loop | 24V DC / 24-mA | |
HART mA loop resistance | 250 Ω ± 20% | |
Special features | Step and ramp function Automatic and manual mode √x, x2: For the measuring function | |
Continuity test | Adjustable threshold up to 100 Ω | |
Power supply | 3.7V / 2300-mAh Li-ion battery | |
Charging time | About 5 h | |
Power adapter | Input: 100 ... 240V AC / 50/60 Hz Output: 5V / 1 A DC | |
Battery life | Approx. 18 h: Simulation and measurement with low LCD illumination, approx. 8 h: Measurement with low LCD illumination | |
Dimensions | 162 x 82 x 40 mm / 6.4 x 3.2 x 1.6 in | |
Weight | About 300 g / < 1 lb | |
Degree of protection | IP20 | |
Operating conditions | Battery operation: 0 ... 55°C / 32 ... 131°F, 30 ... 90% RH Main operation: 0 ... 45°C / 32 ... 113°F, 30 ... 90% RH | |
Storage conditions | -20 .. 60°C / -4 ... 140°F, 30 ... 90% RH non-condensing | |
Heating time | About 15 minutes |
Delivery Scope:
1 x Process Calibrator PCE-LOC 20
2 x Laboratory cables with 2 mm plug
2 x Adapter 2 mm on 4 mm laboratory cable
2 x Alligator clips
1 x Mini USB cable
1 x Power supply 5V / 1 A
1 x Carrying bag
1 x Software